 Dual Attenuation 
Phantom 
Model 406 LE   | 
  |  
| 
   
  
The Gammex 406 LE Dual Attenuation Phantom gives you the benefit of having two phantoms in one with background 
attenuations of 0.5 and 0.7 dB/cm/MHz in a side-by-side configuration. Quality control tests over a wide range of  
frequencies can be performed quickly and easily, providing a comprehensive profile of the scanner¡¯s overall image quality ¨C 
all with a single phantom.  
  
The 406 LE phantom is a highly effective instrument for demonstrating superior image quality while challenging high 
performance ultrasound systems. The phantom incorporates our Tissue Mimicking gel which provides a smoother 
background texture, and a composite film scanning surface that has improved transmission properties so more of the 
ultrasonic beam can be transmitted and received.   
  
With extended target depths to 16 cm, this phantom provides greater scanner and transducer performance tests of  
resolution, depth of penetration and electronic caliper measurements.  Extra pin targets are ideal for testing high frequency 
transducers which are designed to image the first few centimeters of human tissue. Resolution patterns and all vertical and 
horizontal targets are constructed of 0.1 mm nylon fiber. Three sets of axial resolution targets at 3, 8 and 14 cm, and spaced 
at 2.0, 1.0, 0.5 and 0.25 mm, make this phantom extremely effective for demonstrating high resolution detail and challenge 
the ultrasound system¡¯s resolution capability. Scatter-free cylinders of 2, 4 and 6 mm diameter are embedded in the 
phantom representing blood vessels in human tissue.   
  
Optional accessories for the Gammex line of Ultrasound phantoms include Soft Foam-Lined Carrying Case, Rigid Case, or 
the Precision Ultrasound Transducer Guide. 
 |  
| 
   
 
  |  
| 
 This ultrasound image demonstrates the 0.5 (left) 
and the 0.7 (right) dB/cm/MHz attenuation phantoms 
contained in the Gammex 406 LE Dual Attenuation phantom.  |  
| 
   
 
| 
 SPECIFICATIONS  |  
| 
 High Resolution Tissue Mimicking Gel 
Speed of sound . . 1540 ¡À10 m/s at 22¡ãC 
Attenuation . . . . 0.5 and 0.7 ¡À0.05 dB/cm/MHz 
Targets Anechoic Cysts 
Diameters . . . . . 2, 4 and 6 mm 
Speed of sound . . 1540 ¡À10 m/s at 22¡ãC 
Attenuation . . . . 0.05 ¡À0.01 dB/cm/MHz 
Pin Targets 
Diameter of nylon lines . . . . . 0.1 mm (0.004 in) 
Vertical spacing. . 10 mm at 2 to 4 cm deep; 20 mm at 4 to 16 cm deep 
Horizontal spacing . . . . . . . 30 mm at 2 and 12 cm deep; additional pins spaced at 
                                               20 mm in 2 cm set 
Axial resolution . . 3, 3 and 14 cm 
Construction 
Scanning surface . . . . . . . Composite Film 
Walls . . . . . . . . . Extruded ABS 
Dimensions . . . . . 23.2x8.25x18.5 cm (9.25x3.25 7.25 in) 
Weight. . . . . . . . 2.8 kg (6 lbs 5 oz) 
All acoustic measurements at 4.5 MHz, 22¡ãC  |   
   |      | 
 
 
 