Function 1. Numerical calculation: calculate systolic blood pressure, mean pressure, diastolic blood pressure, and pulse rate. 2. Automatic measurement mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90 minutes. 3. Manual measurement mode: the user needed to manually start a measurement. 4. Calibration mode: the value of continuity in the provision of cuff pressure 5. Measurement unit: mmHg 6. Serial Communication: RS232 port, TTL-level, 4800 baud rate 7. Self checking: self-checking function available after reset
Adult: 30 to 255 mmHg systolic blood pressure An average of 20 to 235 mmHg pressure 15 to 220 mmHg diastolic blood pressure Pulse rate of 30 to 240 times / min
Newborn: 30 to 135 mmHg systolic blood pressure An average of 20 to 125 mmHg pressure 15 to 110 mmHg diastolic blood pressure Pulse Rate of 40 to 240 times / min
Measurement time
Adults: up to more than 120 seconds; Newborn: up to more than 90 seconds;
Restriction on software over-voltage
Adult: Less than 290mmHg; Newborns: no more than 145mmHg;
Protection on hardware over-voltage
Adult: 300 ±10 mmHg; Newborn: 150 ±4 mmHg;
Measuring principle
Oscillometry
Deflation approach
ladder Deflated
The initial inflation pressure value
Adult: 150 mmHg; Newborn: 70 mmHg;
Sensor zero reset
Prior to each measurement
Resolution
1 mmHg
Measurement accuracy
Maximum average deviation: no more than ± 5 mmHg; maximum standard deviation: less than 8 mmHg;
Pressure accuracy instructions
± 3 mmHg
Leakage
less than 6 mmHg Per minute;
Electrical performance
Voltage: 12 VDC (adapt to the corresponding valve), 5 VDC Quiescent current: 80 mA Measurement current: 200 mA inflated current: 600mA
Environment
Operating Temperature: 0 to 55 Celsius degrees Storage Temperature: -20 to 65 Celsius degrees HumidityRange: 0 to 95%, non-condenser Operation Altitude: -500 m to 4600 meters Life Span: can operate 100,000 blood pressure measurements
Size
12.6 cm x 6.8 cm x 2.8 cm
Weight:
180 g
Characteristics:
Small in size, simple, excellent reliability and ease of installation clinical monitoring of the accuracy; Excellent reliability and accuracy of clinical monitoring
The latest digital signal processing technology to achieve anti-noise and software algorithms anti-interference by moving arm; Capable of accomplishing all NIBP measurement items; Comply with CE safety design; Comply with all NIBP standard of ANSI/AAMI; Applies to adults, children and Newborn; 5V/12V single powersupply to avoid interference with the digital circuit by the pump; Fully compatible with the protocol of CAS Non-invasive blood pressure module (NIBP)
NIBP Module Model NB001, Small, simple, excellent reliability and ease of installation clinical monitoring of the accuracy with the latest digital signal processing technology to achieve anti-noise and software algorithms anti-interference by moving arm, capable of accomplishing all NIBP measurement items, comply with CE safety design. The evaluation Kit includes the following items: 1) The NB001 module;
2) Lemo connector(male);
3) Lemo connector(female);
4) Extended cannula;
5) 2-way cannula commutator;
6) Cuff;
7) Cannula in instrument;
8) Power cable;
9) RS232 cable Click for More Details