- Automatic heating,infusion at constant temperature 
- Large size multi color LCD screen 
- All data synchronized display, easy to operate 
- Door-free structure 
- Alarm functions 
- Compatible with IV sets of any Brand with high accuracy
 
Performance Parameter
- Standard mode infusion rate: 1mL/h~1300mL/h 
- Max. volume limit : 9999mL (minimum increment 1mL) 
- Max. volume accumulated:  9999mL  (minimum increment 0.01mL) 
- BOLUS and PRIME rate : 600mL/h, 700mL/h, 800mL/h, 900mL/h, 1000mL/h 
- KVO rate: OFF or 1.0mL/h~5.0mL/h (auto mode) 
- Accuracy: ��5% 
- Occlusion alarm threshold: 300mmHg, 500mmHg, 700mmHg, 900mmHg, 1100mmHg 
- Air Bubble Detect: not less than 30��10-3mL 
- Pump Working Noise:  Without Alarms, IP300��s working noise should be under 45dB (A). 
- Alarm Volume: Alarm volume should be over 65dB (A) during alarms. 
- Battery Capability:  Infusion Pump can work under intermittent rate (25ml/h) longer than 4 hours (at 20��C)
 
Alarm
Failure, Low battery, AC failure, Door open, Occlusion, Air in line, Infusion finished (Auto switch to KVO)
 
Power supply
- AC power: AC100 to 240V, 50Hz/60Hz  
- Battery: NiMH 9.6V  
- Power consumption: ��30 VA
 
Physical specification
- Dimensions: 240mm(H)��110mm(W)��230mm(D) 
- Weight:  ��3.5 kg (packaging-free) 
- Screen: Large multi color LCD
 
Environment requirements
- Temperature Operation: +5��C~+40��C (special), +15��C ~+40��C (standard) 
- Storage: -20��C ~+55��C 
- Relative humidity Operation: 20%~90% 
- Storage: ��93%  
- Atmospheric Pressure Operation: 70~106kPa 
- Storage: 50~106kPa 
- Altitude Operation: 500~800mmHg (3565 ~ -440m)  
- Storage: 375~800mmHg (5860 ~ -440m) 
 
 
 There are no documents download yet.