The MiniSim also simulates 45 different arrhythmias including atrial, ventricular, tall T wave, ST elevation, ST depression, myocardial infraction, blocks, pulse, and pacer waveforms. In addition, two invasive blood pressures are simulated with choice of 12 static pressures and 6 dynamic waveforms. Respiration is simulated with 4 rates. 4 base line impedance, and 4 delta impedance selections. Apnea may be off, continuous, and with either 12 or 32 second apnea bradycardia. Finally, temperature is available at either 25 or 37 degrees Centigrade.
The MiniSim 1000 multi-parameter patient simulator is powered by a 9 volt alkaline battery with a typical life of 40 continuos hours. The instrument provides unmatched choices of parameter simulation in a truly compact instrument.
Feautures
|
|
Easy to Use |
|
Small, Compact Size |
|
Menu driven operatoin |
|
Large LCD Display |
|
Tactile Keypad |
|
Battery or AC Operation | |
ECG: 12 leads with independent outputs referenced to RL.
NORMAL SINUS RHYTHM (NSR)
ECG Rates: 30, 60, 70, 80, 90, 100, 120, 150, 180, 210, 240, 270, 300, and 350 BPM. Accuracy 0.5%.
Amplitudes: 0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV on Lead II. Lead 1 is 0.6 X Lead II, Lead III is 0.4 X Lead II, and Lead V is the same as Lead II.
High Level: 500 X low level output on Lead II.
Accuracy: 2% (1-5mV).
PERFORMANCE WAVEFORMS (PERF)
Sine, Square, Triangle, Pulse
Frequencies: 0.1 to 0.9 in 0.1 Hz steps.
1.0 to 9.0 in 1.0 Hz steps.
10 to 100 in 10 Hz steps.
Accuracy: 1%
Amplitudes: 0.1, 0.2, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV on Lead II.
Pulse: 20 ms pulse of 1mV amplitude repeated at 4 second intervals.
R WAVE DETECTION (RWD)
Widths: 10, 40, 50, 60, 70, 80, 90, 100, 110, and 120 ms.
Amplitudes: 0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, 5.0 mV on Lead II.
PACER (PCR)
AP: Atrial Pacer set at 70 BPM.
ASP: Asynchronous Pacer � Ineffective Pacing.
Pacer Amplitudes: + 2 mV and �2mV.
QRS Amplitudes: 1 mV
PCR: Pacer pulses alone.
Pacer Widths: 0.1, 0.2, 0.5, 1.0, 1.5, and 2.0 ms.
Amplitudes: 0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV.
VP: Ventricular Pacer set at 70 BPM.
AVP: Atrial Ventricular Pacer set at 70 BPM.
QRS Amplitude: 1 mV.
Pacer Amplitude: -2mV.
RESPIRATION
Baseline Impedance: 250, 500, 750, and 1000 Ohms.
Delta Impedance: 0.1, 0.5, 1.0, and 1.5 Ohms.
Accuracy: 10%.
Rates: 15, 30, 60, and 120 BPM.
Accuracy: 1%
Apnea: Off, continuous, 12 seconds and 32 seconds.
Lead Configuration: Leads I and II.
BLOOD PRESSURE
Impedance: 350 Ohms.
Excitation: 2 to 16 Volts.
Sensitivity: 5mV/V/mmHg.
Static: 0, 5, 10, 20, 25, 30, 40, 50, 100, 150, 200, and 300 mmHg.
Dynamic: 50/10, 60/20, 70/30, 80/40, 100/60, and 120/80 mmHg. Dynamic waveforms track all NSR rates.
Accuracy: 1%
TEMPERATURE
Compatibility: YSI 400 and 700 Series.
Temperature: 25 and 37 degrees Centigrade.
Accuracy: 2% of setting
POWER REQUIRMENTS:
One 9 Volt alkaline battery or optional AC adapter.
PHYSICAL CHARACTERISTICS:
Dimensions: 5.5 X 3.5 X 1.5 inches (13.9 X 8.9 X 3.8 cm).
Weight: 10 oz ( 0.3 kg).
TEMPERATURE REQUIREMENTS:
Operating: 59 to 95 Degrees F (15 to 35 Degrees C).
Storage: 32 to 131 Degrees F (0 to 55 Degrees C).