The IBP Module Model B300X, the external version, uses PLUG & PLAY IBP TECHNOLOGY with dual-channel measurement. The IBP Module also features low-power, no extra power supply is needed, automatic digital data processing, repeatedly zeroing, calibration function, and can be used to monitor ART, PAP, CVP, RAP, LAP, ICP and P1-P2
ART: Arterial pressure pressure
PAP: pulmonary artery pressure
CVP: central venous pressure
RAP: Left atrial pressure
LAP: right atrial pressure
ICP: intracranial pressure
P1-P2: Expansion of the pressure
Main Board Dimension |
53.1 x 40.6 x 20mm |
Probe Board Dimension |
45.2 x 20.6 x 10mm |
Number of Channels |
Dual Channel |
Pressure Name |
ART, PAP, CVP, RAP, LAP, ICP, P1, P2 |
Range of measurement And alarm |
ART 0 - 300 (mmHg) PA -6 - 120 (mmHg) CVP/RAP/LAP/ICP -10 - 40 (mmHg) P1/P2 -10 - 300 (mmHg) |
Zero-time |
Every time upon measurement |
Calibration |
Regularly or when new sensor is used |
Pressure Sensor |
Sensitivity |
5µV/V/mmHg |
Impedance range |
300 - 3000 (Ohm) |
Resolution |
1 (mmHg) |
Accuracy |
±2% or ±1mmHg, whichever is bigger |
Update Time |
About 1 second |