SPECIFICATION MATRIX of ETCO2 Module MODEL C002A, Respironics LoFlo™ Sidestream CO2 Sensor and CPT WFA Sidestream ETCO2 module


SPECIFICATION MATRIX

ETCO2 Module MODEL C002A, Respironics LoFlo™ Sidestream CO2 Sensor and CPT WFA Sidestream ETCO2 module

 

Items

GoldWEI MODEL C002A Sidestream ETCO2 module

Respironics LoFlo™

Sidestream CO2 Sensor

CPT WFA Sidestream ETCO2 module

Price

$450

$630

$750

Transducer Type/ Sample Rate

Sidestream CO2 100ml/min(default), 50-200ml/min can be configured

Sidestream CO2 50ml/min

Sidestream CO2 100ml/min(default), 50-200ml/min can be configed

Principle of Operation

 

Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts.

Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts.

Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts.

Initialization Time

 

Capnogram displayed in less than 5 seconds, at an ambient temperature of 25°C, full specifications within 2 minutes.

Capnogram displayed in less than 20 seconds, at an ambient temperature of 25°C, full specifications within 2 minutes.

Capnogram displayed in less than 5 seconds, at an ambient temperature of 25°C, full specifications within 2 minutes.

CO2 Measurement Range

020

020

013

CO2 Accuracy

 

±2mmHg  0 - 40 mmHg

±5% of reading 41 - 150 mmHg

±2 mmHg 0 - 40 mmHg

±5% of reading 41 - 76 mmHg

±10% of reading 77 - 99 mmHg

± 2 mmHg0 - 5.0% CO2

< 10% of reading> 5.0% CO2

Respiration Rate Range

2150BPM

2150BPM

3150 BPM

Temperature and Humidity

 

Operating: 0 to 55ºC, 10 to 90% RH, non-condensing

Storage: -40 to 70ºC, <90% RH, non-condensing

Operating: 0 to 40ºC, 10 to 90% RH, non-condensing

Storage: -40 to 70ºC, <90% RH, non-condensing

Operating: 0 to 50ºC, 10 to 90% RH, non-condensing

Storage: -40 to 70ºC, <90% RH, non-condensing

Baro pressure auto compensations

Enable

Unviable, need the host to supply baro pressure value

Enable

Auto offset calibration

Enable

Unviable

Enable

IR source

Long life IR source, 100,000hrs

Custom-built IR source, 20000hrs

Common IR source, 20000hrs

IR detector

Thermopile, which can eliminate Interference of temperature fast changing

Custom-built IR detector, which cannot eliminate Interference of temperature fast changing

Thermopile, which can eliminate Interference of temperature fast changing

Voltage Requirements

 

5.00 VDC (± 5%)

480mW (typical with pump on)

1200mW (extreme status)

5.0 VDC ± 5%

160 mA (800 mW) typical,

290 mA (1.45 W) max.

Ripple not to exceed < 50 mV rms.

5.00 VDC (± 5%)

1000mW (typical with pump on)

Communication interface

TTL / RS232 Optional

RS232 Only

TTL Only

Warranty

2 years

1 year

1year

Custom Requirements

Supply custom-built module @ more than 500 modules per order

Unavailable

Unavailable

 


Model SP100 SPO2 Module (PN: MO-SP100)
Model SP100 SPO2 Module (PN: MO-SP100)
Patient Monitor MGW6012
Patient Monitor MGW6012
NIBP Module  Model NB001 (MONM-BN001)
NIBP Module Model NB001 (MONM-BN001)
SmartSat Pulse Oximetry Analyzer
SmartSat Pulse Oximetry Analyzer

About Us | Latest News | Products| Category List | Terms & Warranty | Manufacturers | Knowledge Base | FAQ | Contact Us |JJWEI Computers | GoldWEI Superfood |
© 2003-2011, GoldWEI Corporation, All Rights Reserved.  Phone: 703-713-0198  Email: sales#goldwei.com